Part Number Hot Search : 
IRF132 596AC MAX1667 HMC128G8 M71C4 C2562 820000 M38197
Product Description
Full Text Search
 

To Download 1550A Datasheet File

  If you can't view the Datasheet, Please click here to try to view without PDF Reader .  
 
 


  Datasheet File OCR Text:
 1550A
Vishay Micro-Measurements
Strain Indicator Calibrator
FEATURES
* True Wheatstone bridge circuitry * Simulates quarter, half & full bridge -- both 120/350 * Three decades of push buttons * Strain range direct reading: 99 900e. . . increments of 100e * Transducer range: 49.95 mV/V. . . increments of 0.05 mV/V * Reversing switch for plus and minus calibration * High precision Vishay resistors used throughout to ensure excellent stability * Accuracy 0.025 percent -- traceable to the U.S. National Institute of Standards and Technology
A laboratory standard for verifying the calibration of strain and transducer indicators.
DESCRIPTION
Sound engineering and laboratory practices require that the instrumentation used to make critical strain measurements be periodically calibrated to verify that it is within the manufacturer's original specifications. Additionally, each type of strain indicator exhibits some degree of nonlinearity, especially for large strains during quarter-bridge operation. Since this is the most common stress analysis application of strain gages, it is important that the strain indicator be calibrated in this mode. Instrumentation span should also be checked at a number of points before each important test to avoid inaccurate data. The Model 1550A calibrator is a Wheatstone bridge and generates a true change of resistance in one or two arms of the bridge. It simulates the actual behavior of a strain gage in both positive and negative strain. The `star network' used in certain other commercial calibrators provides a substantially lower cost instrument design, because component specifications are less critical, and fewer components are required. However, the `star network' cannot simulate quarter-bridge strain gage behavior, and cannot simulate positive strain. Another serious problem with this circuit is that the bridge input and output resistances change in an abnormal manner, leading to inaccuracies in calibration under some conditions. A calibrator based on the Wheatstone bridge principle requires stable components. A total of 66 ultra-stable Vishay precision resistors are used in the Model 1550A calibrator to provide the stability, repeatability, accuracy and incremental steps required in a laboratory standards instrument.
WHEATSTONE BRIDGE / STAR NETWORK
Document Number: 11313 Revision 03-Feb-03
micro-measurements@vishay.com
www.vishaymg.com 1
1550A
Vishay Micro-Measurements
Strain Indicator Calibrator
A certificate of calibration is provided with each Model 1550A calibrator
SPECIFICATIONS
Accuracy: 0.025% of setting 1 (0.0005 mV/V), maximum. Traceable to United States National Institute of Standards and Technology. Repeatability: 1 (0.0005 mV/V), maximum Stability: (0.001% of setting 1)/C, maximum Thermal EMF: 0.5 V/V of excitation, maximum Bridge Resistances: 120 and 350. Input Resistance: 0.05%, maximum, from nominal at all output settings. Output Resistance: 0.05%, maximum, from nominal at "000", -0.25% at 99 900. Circuit: True R in two adjacent arms (opposite signs), plus two fixed arms for bridge completion. Simulation: Quarter bridge, one active arm. Half bridge, one or two active arms. Full bridge, two active arms. Range: Two active arms: 0 to 99 900 in steps of 100 @GF = 2.00. 0 to 49.95 mV/V in steps of 0.05 mV/V. One active arm: 0 to 49 950 in steps of 50 @ GF = 2.00. Excitation: To meet accuracy and repeatability specifications: 120: Up to 10 Vdc. 350: Up to 15 Vdc. Maximum Permissible: 120: 25V ac or dc. 350: 30V ac or dc. Output @ 000: 50 (0.025 mV/V), maximum in full-bridge mode. Environment: Temperature: +50 to +100F (+10 to +38C). Humidity: Up to 70% RH, non-condensing. Size: Aluminum case (separable lid): 5-3/4 H x 8-1/4 W x 7-3/4 D in (145 x 210 x 195 mm). Weight: 4.8 lb (2.2 kg). All specifications are nominal or typical at +23C unless noted.
www.vishaymg.com 2
micro-measurements@vishay.com
Document Number: 11313 Revision 03-Feb-03